SV3000 Saybolt Viscosity Bath
Catalog/Part Number:: SV3000
Description
Test
Method
Determines
the time required for 60mL of sample to flow through a calibrated orifice under
precisely controlled conditions. Saybolt Universal Seconds (SUS) is the
standard measurement for lubricants, insulating oils and lighter fuel grades,
and Saybolt Furol Seconds (SFS) is used for heavier oils and bitumens.
Features
Microprocessor
control of temperature between ambient and 240 °C (464°F)
Four tube
capacity
Dual digital
displays show setpoint and actual temperature
Selectable
temperature scale - Celsius or Fahrenheit
Conforms to
ASTM D88, D244, E102, and related specifications
Description
Constant
temperature bath for viscosity determinations using Saybolt viscometer tubes
and orifices. Microprocessor PID circuitry assures precise temperature control
within ASTM specified tolerances throughout the operating range of the bath.
Simple push-button controls and dual digital displays permit easy setting and
monitoring of bath temperature. Two place calibration offset is provided.
Accommodates four viscometers and four 60mL receiving flasks. Sliding draft
shields and a chemical-resistant alignment plate facilitate handling of the
flasks, and glare-free fluorescent backlighting is provided for easy viewing of
the samples. Insulated bath interior is constructed entirely of heavy gauge
stainless steel. A built-in overflow pipe and drain valve simplifies filling of
the bath fluid to the proper level. Chemical resistant top plate provides
excellent insulation and is easily removed to allow for cleaning of the bath
interior. A cooling coil for tap water or refrigerated coolant is provided for
operation at near-ambient temperatures. Steel cabinet has leveling feet and a
chemical resistant polyurethane-epoxy finish.
Included
Accessories
Cleaning
Plunger
Oil Strainer
Tube Nut
Wrench
Port
Closures
Chained
Corks
Withdrawal
Tube
Orifice
Wrench
Port Covers
Thermometer
Supports
Technical Specifications
Specifications
Conforms to
the specifications of: ASTM D88, D244, E102, AASHTO T72; FTM 791-304
Capacity: 4
viscometer tubes
Temperature
Range: ambient to 464 °F (240 °C)
Temperature
Stability: ± 0.05 °F (± 0.03 °C)
Bath
Capacity: 5 gal (19 L)
Recommended
Bath Medium: water or suitable heat transfer fluid
Electrical
Requirements
220-240V
50/60Hz, Single Phase, 6.4A
Additional
Accessories
K22009
Saybolt Viscometer Tube, Brass
K22309
Saybolt Viscometer Tube, Stainless Steel
K22010
Saybolt Universal Orifice
K22010-C/F
Saybolt Universal Orifice with Calibration Certificate
K22020 Furol
Orifice
K22020-C/F
Sayblt Furol Orifice with Calibration Certificate
K22029
Kansas Road Oil Orifice
332-003-003
Borosilicate Glass Receiving Flask, 60 mL, for SV3000
K22030
Orifice Wrench for Universal and Furol Orifices
K22039
Orifice Wrench for Kansas Road Oil Orifice
K22050
Socket Wrench
K22060 Oil
Strainer
K22070
Cleaning Plunger
K22080
Displacement Ring
K22090
Withdrawal Tube
K22011
Thermometer Support
355-001-001
White Technical Oil, 1 Gallon Container
355-001-002
Silicone Heat Transfer Fluid, 1 Gallon Container
355-001-003
White Technical Oil, 5 Gallon Container
355-001-004
Silicone Heat Transfer Fluid, 5 Gallon Container
250-000-17F
ASTM 17F Thermometer
250-000-17C
ASTM 17C Thermometer
Zero(0) Reviews