Four Ball Tester
Catalog/Part Number:: K93100
Description
Tesrt
Method
Determines
the Wear Preventative (WP) and Extreme Pressure (EP) characteristics of
lubricating oils and greases in sliding steel-on-steel applications. The test
consists of rotating a steel ball under load against three stationary steel
balls coated with lubricant. Measurements are taken at the rotating speeds,
temperatures, and duration as specified by published standards. The load-wear
index can be calculated from the weld point in EP tests, and lubricant
comparisons can be made based upon scar diameters incurred from wear tests.
Features
- Conforms
to ASTM D2266, D2596, IP 239, and related specifications
- Performs
Wear Preventative (WP) and Extreme Pressure (EP) tests
- Displays
and records normal load, frictional torque, time, and temperature
- Test
speeds and temperatures are electronically controlled
- Data
Acquisition Software and Card are included
- Custom
configurations are available
- Precise
variable loading capability
Description
Four
Ball Tester performs both Wear Preventative (WP) and Extreme Pressure (EP) analyses for measuring the wear and frictional properties of
lubricants under sliding steel-on-steel
test conditions. Tests are performed in accordance to the latest ASTM and IP published methods. Normal load on the ball assembly and frictional torque are measured through load cells. Data is processed and stored using an advanced data acquisition and processing software package. Test results can be plotted and compared, as well as exported to other programs. Wear scars on the steel balls are measured and recorded with a High Resolution Digital Microscope available
as recommended accessory for the Four Ball Tester. High Resolution Digital Microscope Koehler's Four Ball Microscope is a versatile device for measuring the
wear scar diameter on a steel test ball. This apparatus consists of the
“Dinolite” Microscope with “DinoCapture”
Software mounted at an angle on an aluminum base. The device is designed to measure the wear scar without removing the test balls from the ball pot allowing for a safer measurement procedure. The wear scar can be viewed through an external PC. The
software measures the wear scar using a
diameter and line tool. The images can be saved at varied resolutions on a PC.
Included Accessories
- Ball
Chuck Remover
- Ball
Rack
- Ball
Clamp Ring
- Ball
Holder Base Disc
- Set
of Hand Tools
- Torque
Wrench
- Electrical
Controller
- Connecting
Cables
- Data
Aquisition Software
- Calibration
and Test Reports
- Ordering
Electrical Requirements
440V, 50Hz, 3 phase
220V,
50Hz, 3 phase
Dimmenssions
38x48x52 in. (96.5x121.9x132.1 cm)
Additional Accessories
- K93105
Test Balls (Pack of 100)
- K93111
High Resolution Digital Microscope
Technical Specifications
Conforms to the specifications of: ASTM
D2266, D2596, D2783, D4172, D5183*, IP 239
Drive Motor: 1.5 kW
Test Speeds: 1200, 1440, 1760 rpm
Optional Test Speeds (min/max): 1000/3000,
300/3000 rpm
Maximum Axial Load: 10000 N at 3000 rpm or
12000 N at 1800 rpm
Test Duration (min/max): 1/9999 min
Test Ball diameter: 12.7 mm
Zero(0) Reviews