Digital Penetrometer
Catalog/Part Number:: K95500
Description
Test Method
Penetration tests are performed on petroleum products to
determine consistency and shear stability (lubricating greases) for
design, quality control and identification purposes. A standard cone or
needle is released from a penetrometer and allowed to drop freely into the
sample for 5 seconds (or a different specified interval) at constant
temperature. The depth of penetration of the cone or needle into the
sample is measured in tenths of a millimeter by the penetrometer.
Features
Tests the consistency of lubricating greases, petroleum waxes,
bitumens, pastes, creams and other solid to semi-solid products
Automatically timed operator programmable penetration measurements
Motorized placement of penetrator on sample surface
Large LCD to display all functions
RS232 port for data transfer
Full measurement range of 0-620 in 1⁄10mm scale or 1⁄100mm scale
Rechargeable battery or AC operation
Large, removable base accommodates grease worker cups and other
ASTM and non-standard sample containers
Complete selection of penetrometer cones, needles and accessories
for petroleum products testing and for a wide range of other applications
Conforms to all ASTM, IP, ISO 9001 and related specifications for
penetrometers
Description
Microprocessor based penetrometer loaded with advanced features to
provide ease of operation and highly reproducible consistency measurements of
petroleum products. Microprocessor control provides a full range of measurement
and reporting options, and operation is simplified by four user programmable
presets that facilitate lowering the penetrator tip to the sample surface.
Automatically timed penetrations—The penetrometer defaults to the
standard ASTM interval of 5.0 seconds, or the operator may conveniently program
a different interval in the range between 0.1 and 9999.9 seconds (in 0.1 second
increments). A curing or temperature stabilization period may also be programmed
by the operator (to delay the release of the penetrator into the sample) and
for added convenience all selected parameters are retained in memory and
automatically repeated in subsequent tests until changed by the operator.
Separate keypad controls for each parameter simplify operation. Penetration and
delay intervals count down on a large, easy to read LCD on the head of the
unit.
Convenient measurement and reporting options—Penetration
measurements in the full range of 0 - 620 in 1⁄10mm scale are reported in
either 1⁄10mm or 1⁄100mm increments at the operator’s option. For quality
control testing, a penetration range can be entered into memory prior to
testing. If a test result falls outside of the programmed range, an audible
signal and visual error message alert the operator of a failed sample. Test
results are displayed in digital format on a large LCD readout on the head of
the penetrometer and can be communicated to a printer or computer via a
built-in RS232 interface.
Simplified penetrator tip placement—Correct placement of the
penetrator tip on the sample surface is essential for accurate penetration test
results. The Koehler Digital Penetrometer has four operator programmable
presets that lower the penetrator to the sample surface height at the touch of
a button, greatly simplifying the process to ensure reproducibility. A fine
adjustment button permits slight adjustments as needed. Full manual operation
is also available with the use of coarse and fine push button controls and
built-in magnifier and illuminator arms. When testing electrically conductive
samples, a built-in circuit senses the sample surface for automatic placement.
After testing, the penetrometer head returns to a raised position at the touch
of a button to facilitate cleaning of the penetrator and changing of the
sample.
More convenience features—The detachable machined base provides a
large platform to accommodate a wide range of sample containers and constant
temperature cylinders. It removes easily to permit the head assembly to be
reversed (for use with a constant temperature bath) or mounted directly to a
bath housing or other location. A built-in rechargeable battery pack permits
field operation and provides back-up in the event of power interruption.
Battery recharges automatically during operation of the penetrometer on
standard AC electrical service.
Included Accessories
Standard Plunger, 47.5g
Weights, 50 and 100g
Technical Specifications
Specifications
Conforms to the specifications of: ASTM D5, D217, D937, D1321,
D1403,
D2884, D4950, D5329; IP 49, 50, 179, 310; ISO 2137; DIN 51579,
51580, 51804; FTM 791-311, 791-312, 791-313; AOCS Cc 16-60; AACC 58-14; NFT
60-119, 60-123, 60-132, 66-004
Penetration Range: 0-62.0mm (0-620 penetration scale) in 1⁄10mm or
1⁄100mm
Penetration Interval: Operator variable from 0.1 to 9999.9 seconds
with automatic repeat function and 5.0 second default
Electrical Requirements
115V 60Hz
220-240V 50/60Hz
Dimensions
lxwxh,in.(cm)
Base: 12 1/2 x 14 (31.7x35.6)
Overall: 12 1/2 x 14 x 18 (31.7x35.6x45.7)
Net Weight: 21 lbs (9.6 kg)