K27580 Automatic Petroleum Refractometer w/ Touch Screen

Availablity: EXW

Catalog/Part Number: K27580
Brand: KOEHLER
Supplier: KOEHLER
Supplier Location: Holtsville, New York 11742, USA
Inventory Weight: unavailable
Unit of Measurement:
Inventory Brochure: unavailable

Catalog/Part Number:: K27580

Sorry no inventory found for this brand at this time.

Description

Test Method

Refractive Index is a fundamental physical property that is used in conjunction with other properties to characterize pure hydrocarbons and their mixtures. It is a useful property for concentration measurements, purity determinations and chemical identification.

 

Features

Conforms to ASTM D1218, D1747, D5006, and D5775 test specifications

Electronic heating and cooling twin Peltier system operates from 10 to 100°C, eliminating the need for a circulating water bath

Automated and precise refractive index measurements

Detection of correct sample loading with warning message

Monitors cleaning between samples and warns if it the instrument is not cleaned correctly.

Designed for samples ranging from clear to highly colored, dark and opaque

Measures to the fifth decimal place refractive index or hundredths place in percent solids

Multipoint calibration routines maximize accuracy

Automatic detection of temperature and reading stability

Full networking capability with direct data transfer into

Excel, .pdf and other formats

 

Description

The Koehler Automatic Refractometer uses precision optics and superior image analysis to extend the repeatability and accuracy of refractive index measurements for petroleum products. Subjectivity is removed from tests results because no manual activities such as aligning shadowlines or reading analog scales are necessary. Opaque hydrocarbons present no problem for this unit which uses reflected light measurement technology as opposed to manual refractometers which are of the transmission type. The refractometer incorporates numerous innovations designed to improve the accuracy of petroleum product testing. A linear scanned diode array shadowline sensor is sensitive enough to read even the most demanding samples. A 589 nanometer filter gives true Sodium D-Line refractive index readings. A built-in twin Peltier heating and cooling system gives the instrument the capability of reading at up to 100°C without the need for an accessory water bath. The measuring system guides the operator through every step of the process, advising if enough sample has be used and if it’s placed correctly, providing information about the cleaning process and if residue has been left behind. Set-up, diagnostic and calibration routines are displayed with easy to follow step-by-step instructions. User-developed customer calibration curves may be programmed allowing automatic temperature correction, direct percent concentration, percent reaction completion, etc. This unit has been used successfully throughout the petrochemical industry.

" autofocus>

Test Method

Refractive Index is a fundamental physical property that is used in conjunction with other properties to characterize pure hydrocarbons and their mixtures. It is a useful property for concentration measurements, purity determinations and chemical identification.

 

Features

Conforms to ASTM D1218, D1747, D5006, and D5775 test specifications

Electronic heating and cooling twin Peltier system operates from 10 to 100°C, eliminating the need for a circulating water bath

Automated and precise refractive index measurements

Detection of correct sample loading with warning message

Monitors cleaning between samples and warns if it the instrument is not cleaned correctly.

Designed for samples ranging from clear to highly colored, dark and opaque

Measures to the fifth decimal place refractive index or hundredths place in percent solids

Multipoint calibration routines maximize accuracy

Automatic detection of temperature and reading stability

Full networking capability with direct data transfer into

Excel, .pdf and other formats

 

Description

The Koehler Automatic Refractometer uses precision optics and superior image analysis to extend the repeatability and accuracy of refractive index measurements for petroleum products. Subjectivity is removed from tests results because no manual activities such as aligning shadowlines or reading analog scales are necessary. Opaque hydrocarbons present no problem for this unit which uses reflected light measurement technology as opposed to manual refractometers which are of the transmission type. The refractometer incorporates numerous innovations designed to improve the accuracy of petroleum product testing. A linear scanned diode array shadowline sensor is sensitive enough to read even the most demanding samples. A 589 nanometer filter gives true Sodium D-Line refractive index readings. A built-in twin Peltier heating and cooling system gives the instrument the capability of reading at up to 100°C without the need for an accessory water bath. The measuring system guides the operator through every step of the process, advising if enough sample has be used and if it’s placed correctly, providing information about the cleaning process and if residue has been left behind. Set-up, diagnostic and calibration routines are displayed with easy to follow step-by-step instructions. User-developed customer calibration curves may be programmed allowing automatic temperature correction, direct percent concentration, percent reaction completion, etc. This unit has been used successfully throughout the petrochemical industry.

Catalog/Part Number: K27580 Category:

Technical Specifications

Specifications

Conforms to the Specifications of:

ASTM D1218, D1747, D5006, D5775

Measurement Scales: Refractive Index (RI), BRIX (% sucrose), Temperature Corrected RI, Temperature Corrected BRIX, Ten User-Programmable Scales

Illumination: 589nm light emitting diode with interference filter (estimated life: 100,000 hrs)

Range: Dissolved Solids: 0 to 95% solids

Refractive Index: 1.26000 to 1.72000nD (nD - Sodium D-Line Refractive Index)

Readability: 0.01% Solids 0.00001nD

Precision: Refractive Index Standard Oils ± 0.00002 - Typical clear aqueous samples, % Solids Temperature Compensated, as sucrose ±0.02%

Calibration Fluid: refractive index standard oil, NIST traceable nominal value 1.495 RI, 67.61 BRIX

Sample Types: Transparent, translucent or opaque

Prism Assembly: Stainless steel, synthetic sapphire sealed with solvent-resistant epoxy

Calibration: 1 point - Water only;  2 point - Water and refractive index or Brix standard

 

Electrical Requirements

85 to 260 VAC; 48 to 62 Hz

 

Dimensions

lxwxh,in.(cm) : 17 x12 x 13 (44x31x33cm)

 

Additional Accessories

K27505 Calibration Fluid

Certificate of NIST Traceability Included

Add a review

You must login to make reviewLogin

  1. Zero(0) Reviews